Phase Locking using EIP 371
In a pinch, I have used my EIP 371 Source Locking Microwave Counter to phase
lock an otherwise unstable oscillator (Marconi/IFR 6310 2-20 GHz sweeper).
EIP 371 Source Locking Counter
The picture below was obtained by phase locking a Marconi 6310 sweeper
(with an FM input going down to DC). The resolution of the counter's
phase locking circuit is 100 kHz (programmed via front panel keypad).
The loop bandwidth is quite narrow, but in some cases that's all that's
needed, and all you need is a BNC cable from the output of the EIP 371
to the FM control of the VCO. The counter figures the polarity and loop
gain by itself. It's magic!
Without phase locking, the signal would not stay in the window of the
analyzer with that span very long. With phase locking, it is dead stable
and quite usable for a number of narrow band experiments.
Compare to the HP 8673 at the same frequency:
>
At 25 kHz, phase noise is about the same, but that may be the floor
of the spectrum analyzer (Tek 492)